5 • Service
Doc. 065259-05 10/12
77
5.4
Manually Raising the Sampling Head
Follow this procedure if you need to raise the sampling head out of the carousel
manually (for example, to clear a needle jam).
The following tools are needed to complete this procedure: #1 Phillips
screwdriver and 2.5-mm Allen wrench.
1. Turn off the Dionex AS-DVpower and disconnect the power cord. Raise the
top cover.
2. Remove the Dionex AS-DV front cover by gripping the indentations at each
side and pulling the cover straight off toward you.
3. Use a #1 Phillips screwdriver to remove the three flathead screws from the
sampling head tower cover (see
). Lift up the cover to remove it.
Figure 5-10. Removing the Sampling Head Tower Cover
Remove 3 screws
Sampling
Head Tower