Dionex AS-DV Operator’s Manual
38
Doc. 065259-05 10/12
3.7
Starting a Run
After creating the sequence, you can start a run.
•
For Chromeleon 7:
On the Console, click
Start
on the Sequence Control Bar above the injection
•
For Chromeleon 6.8:
On the panel tabset, click the
Sequence Control
tab. Click
Load Sequence
,
open the sequence you created, and then click
Start Batch
.
3.8
Operating Details
3.8.1
Interrupting the Autosampler
•
To interrupt the autosampler when a sequence is running in
Chromeleon 7:
On the Console, click
Stop
on the Sequence Control Bar above the
injection list. Specify when to stop (immediately, after the current
sample, or after the current sequence).
•
To interrupt the autosampler when a sequence is running in
Chromeleon 6.8:
Select
Batch > Stop
on the Chromeleon menu bar. Select when to
stop (immediately, after the current sample, or after the current
sequence).
Before starting a run, always press the
Carousel Release
button to
disengage the carousel and reset vial memory. After loading the vials
in the carousel, press the
Carousel Release
button again to engage
the carousel.
During a run, do not press the Carousel Release button. Doing so
clears all current vial information from Dionex AS-DV memory. (During
a run, the Dionex AS-DV stores the vial size installed in each carousel
position and the remaining volume in each vial.)
Summary of Contents for Dionex AS-DV
Page 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Page 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Page 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Page 132: ...Dionex AS DV Operator s Manual Index 6 Doc 065259 05 10 12 ...