Rotors and accessories
34
Be careful with damaged carbon fiber
rotors!
Never stop a running rotor manually!
Sharp carbon fiber splinters can cause
injuries.
LAC-250
The rotor consists of carbon fiber-epoxide resin com-
pound with metal components.
Compared to traditional metal rotors these composite
rotors show changed characteristics with regard to
chemical resistance. Organic solutions, strong acids
or bases can initiate premature material weakness.
Before use carefully and regularly inspect rotors
made of carbon fiber composite material for surface
wear.
The rotor may not be operated if, due to
damage, fibers escape the rotor surface!
Wear suitable protective gloves (e.g. of
leather) when handling damaged carbon
fiber rotors!
Never expose the rotor to ethylene
oxide!
Doing so will damage the composite
material.