Thermo Scientific Cryofuge 8 / 16 Centrifuges and Heavy Duty Centrifuges
Maintenance and Care
6. Autoclaving
CAUTION
Never exceed the permitted temperature and duration when autoclaving.
NOTICE
No chemical additives are permitted in the steam.
Always disassemble all parts before autoclaving, e.g. lids need to be removed before autoclaving bottles, buckets or rotors.
Within the list of components marked parts can be autoclaved at 121 °C for 20 min.
Make sure that the necessary sterility is achieved according to your requirements.
After autoclaving, treat the entire surface of aluminum parts including the cavities with corrosion protection oil (70009824).
Treat the trunnion bolts of the swing out rotor with bolt grease (75003786).