III–8
Thermo Scientific Cryofuge 8 / 16 Centrifuges and Heavy Duty Centrifuges
Operation
Figure III–12:
Insert windshield lid into holder
CAUTION
Always put the windshield lid into the holder at the inner side of the centrifuge door. Make sure
the rotor is locked properly. A missing windshield lid and an unlocked rotor can lead to severe
damage of rotor and centrifuge.
4. Maximum Loading
Each rotor is designed to run with his maximum load at maximum speed. The safety system of the centrifuge requires that
the rotor is not overloaded.
The rotors are designed to work with substance mixtures with a density of up to 1.2 g/ml. If the admissible maximum load is
exceeded the following steps need to be taken:
Reduce the fill level.
Reduce the speed.
Use the table or the formula:
n
adm
= admissible maximum application speed
n
max
= maximum rated speed
w
max
= maximum rated load
w
app
= applied load