52 | Controller Menu Maps
Thermo Fisher Scientific CryoExtra CE8100 Series
Temperature Calibration Menus
Press ENTER for Temp A
Calibration or Press
SETUP for next menu
Calibration Type
SINGLE POINT
Use
to adjust
Press ENTER for next
Temp A Calibration
Place Probe A in LN
2
and press ENTER
Wait for Temp A to
stabilize then press
ENTER Temp A -
195.8°C
Remove Probe A from
LN
2
and press ENTER
Wait while probe warms
to room temperature
180 seconds
Place Probe A in ice
water and press ENTER
Wait for Temp A to
stabilize then press
ENTER Temp A 0°C
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
Temp A Calibration
Place probe A in
LN
2
and press
ENTER
Wait for Temp A to
stabilize then press
ENTER Temp A -
195.8°C
Probe A Single Point
Calibration Complete
Calibration
aborted
5 sec
time out
ENTER
180 sec
time out
ESC
Probe A Two Point
Calibration Complete
Calibration Type
SINGLE POINT
Use
to adjust
Press ENTER for
next
Temp B Calibration
Place Probe B in LN
2
and press ENTER
Wait for Temp B to
stabilize then press ENTER
Temp A -195.8°C
Remove Probe B from
LN
2
and press ENTER
Wait while probe warms
to room temperature
180 seconds
Place Probe B in ice
water and press ENTER
Wait for Temp B to
stabilize then press
ENTER Temp B 0°C
ENTER
ENTER
ENTER
ENTER
ENTER
Temp B Calibration
Place probe B in
LN
2
and press
ENTER
ENTER
Wait for Temp B to
stabilize then press
ENTER Temp B -
195.8°C
Probe B Single Point
Calibration Complete
180 sec
time out
ENTER
Probe B Two Point
Calibration Complete
ENTER
(TWO)
Calibration
aborted
5 sec
time out
ESC
ENTER
(SINGLE)
ENTER
ENTER
ENTER
(TWO)
ENTER
(SINGLE)
Press ENTER for Temp B
Calibration or Press
SETUP for next menu