Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 83
5.6.2
Connecting Fittings, Capillaries, and Tubing
This section provides information about how to connect and handle
capillaries, fittings, and tubing.
5.6.2.1
General Guidelines
When connecting capillaries and tubing, follow these general
recommendations:
•
Use only the capillaries and tubing (for example, solvent lines or
waste tubing) that are shipped with the detector or additional or
spare capillaries and tubing as recommended by Thermo Fisher
Scientific.
•
The connectors must be free from contaminants. Even minute
particles may cause damage to the system or lead to invalid test
results.
•
Do not install capillaries or tubes that are stressed, nicked, kinked, or
otherwise damaged.
•
Install capillaries and fittings only at the positions for which they are
intended.
TIP
This is especially important in an UltiMate 3000 system, because
different fitting systems are used in the system.
For further information:
♦
See
section 5.6.2.2 Connecting Viper Capillaries, page 84
♦
See
section 5.6.2.3 Guidelines for Conventional Fitting Systems,
page 85
Summary of Contents for Corona Veo
Page 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Page 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Page 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...