1
Rotor Specifications
Technical Data
1-2
75003667 BIOLiner
Thermo Scientific
Technical Data
Table 1-1.
230 V, 50 / 60 Hz, Rectangular buckets 75003668
Centrifuge
Multifuge X3
Multifuge X3 F
Megafuge 40
Catalog #
75004500
75004530
75004503
Weight empty [kg]
7.3
7.3
7.3
Max. cycle number
55000
55000
55000
Maximum permissible load [g]
4 x 750
4 x 750
4 x 750
Maximum speed n
max
[rpm]
3500
3500
3500
Maximum RCF value at n
max
2739
2739
2739
Max. / min. radius [cm]
20 / 10.5
20 / 10.5
20 / 10.5
Pitch angle [°]
90
90
90
Accel. / braking time [s]
55 / 55
55 / 55
55 / 55
Sample heating at n
max
[°C] referred to
ambient temperature of 23-25 °C, running
time 120 minutes
8
8
8
Aerosol-tight
*
yes
yes
yes
Permissible temperature range autoclavable
(cycle number) °C
121
121
121
*
Tested by HPA, Porton-down, UK
Centrifuge
Multifuge X3R
Multifuge X3 FR
Megafuge 40R
Catalog #
75004515
75004536
75004518
Weight empty [kg]
7.3
7.3
7.3
Max. cycle number
55000
55000
55000
Maximum permissible load [g]
4 x 750
4 x 750
4 x 750
Maximum speed n
max
[rpm]
3500
3500
3500
Maximum RCF value at n
max
2739
2739
2739
Max. / min. radius [cm]
20 / 10.5
20 / 10.5
20 / 10.5
Pitch angle [°]
90
90
90
Accel. / braking time [s]
55 / 55
55 / 55
55 / 55
Minimum temperature at n
max
[°C] relating to
ambient temperature 23-25 °C
<0
<0
<0
Speed at 4 °C
3500
3500
3500
Aerosol-tight
*
yes
yes
yes
Permissible temperature range autoclavable
(cycle number) °C
121
121
121
*
Tested by HPA, Porton-down, UK