background image

1

  

Rev 1.0,  Cat. no. N11468 

Thermo Scientific Arktik Thermal Cycler 

 

To install a block

:

 

Step 

Action                                                                                                                                              

Make sure the power is off.   

Raise the handle at the back of the block and hold the block by the front and back edges. Lower the block into the 
base unit, leaving at least 2 cm between the front edge of the block and the front of the base unit. 

Slide the block forward as far as it will go, and push the handle down. Do not use force! The block should click 
into place, and the handle should be completely horizontal. 

 

 

 

 

 
Operation: 

Step 

Action                                                                                                                                              

Connect the power cable and power up the instrument (with the block in place) by switching on the power switch 
on the back of the instrument. 

2 Select 

NEW

 or 

RUN

 from the function keys to load or create a protocol (see the reverse side). 

To access the sample blocks, open the lid by grasping the handle and pulling it forward. 

Place sealed PCR vessels into the block. 

To close the sample block, apply downward pressure on the lid while pulling the handle forward. To tighten the 
lid, turn the dial clockwise. To prevent over-tightening, the lid mechanism will begin ratcheting when the proper 
pressure has been achieved.   

Monitor the run using the 

STATUS

 or 

TIME

 function key. 

7 Press 

the 

CANCEL

 function key to cancel the run or the 

PAUSE

 function key to pause the run. 

 

Once the run completes, press the 

MAIN

 function key to exit to the 

Main

 menu. 

To open the block after a run, first turn the dial counterclockwise to raise the heated lid and then open as in  
Step 3. 

10 

Note that in some cases, a  slight expansion of the plate during thermal cycling may occur, and this may cause 
difficulties in opening the lid. If this occurs, press the release button on top of the dial down while turning the dial 
counterclockwise. 

    

All cross-references refer to the 

Thermo Scientific Arktik 

Thermal Cycler 

User Manual

, Rev 1.0 (Cat. no. 

N11467). 

 

Quick Reference Guide