Prepare samples using MicroAmp

 tube strips with

attached caps

The following procedure describes how to properly load and seal

MicroAmp tube strips with attached caps using the MicroAmp 96-

well tray/retainer set. For a visual demonstration, watch the video

"How to use adapter tray retainers" by visiting 

PCR/qPCR Plastics

and Seals Education

.

1.

Separate the 

blue

 tray from the retainer by squeezing the

release catch as indicated in the graphic.

1

2

3

1

1

Release catch

2

MicroAmp

 96-Well Retainer

3

MicroAmp

 96-Well Tray

2.

Place the 

blue

 tray on the splash-free 96-well base.

Note: 

Only the bottom tray is used for tubes with attached

caps.

3.

Load the tube strips with attached caps into the tray.

4.

Pipette the reaction mixture into the tubes.

5.

Seal the cap strips using the rocking capping tool:

a.

Slip your fingers through the handle with the holes in the tool

facing down for domed caps and with the holes facing up

for optical caps.

b.

Align the tool over the first eight caps in a row.

c.

Rock the tool back and forth a few times to seal the caps.

d.

Repeat for all remaining rows.

6.

Remove the 

blue

 tray containing the sealed tube strips from

the 96-well base and place the tray and sealed tube strips into

the instrument.

1

MicroAmp

 8-Tube Strip with Attached Caps (0.2-mL)

2

MicroAmp

 96-Well Tray

3

MicroAmp

 Splash Free 96-Well Base

Seal tube strips with cap strips

IMPORTANT! 

Apply significant downward pressure on the sealing

tool in all steps to form a complete seal on top of the tubes.

1.

Align and place the cap strips on the tubes.

2.

Seal the cap strips using the rocking capping tool:

a.

Slip your fingers through the handle with the holes in the tool

facing down for domed caps and with holes facing up for

optical caps.

b.

Align the tool over the first eight caps in a row.

c.

Rock the tool back and forth a few times to seal the caps.

d.

Repeat for all remaining rows.

Prepare samples using MicroAmp

 Reaction Tubes

The following procedure describes how to properly load and seal

MicroAmp

 individual tubes with attached caps and the MicroAmp

96-well tray for VeriFlex systems. For a visual demonstration, watch

the video "How to use adapter tray retainers" by visiting 

PCR/qPCR

Plastics and Seals Education

.

1.

Set the 

black

 tray on a 96-well base.

2.

Place the reaction tubes in the 

black

 tray.

1

2

3

1

MicroAmp

 Reaction Tube with Cap (0.2-mL)

2

MicroAmp

 96-Well Tray for VeriFlex

 Blocks

3

MicroAmp

 Splash Free 96-Well Base

3.

Pipette the reaction mixture into the reaction tubes.

4.

Cap the tubes.

5.

Remove the 

black

 tray with sealed reaction tubes from the 96-

well base and place the tray and sealed tubes into the

instrument.

4

VeritiPro

 Thermal Cycler Quick Reference

Summary of Contents for Applied Biosystems VeritiPro

Page 1: ...ro Thermal Cycler IMPORTANT Save the packing materials and box in case you need to ship the instrument in for service 1 Open the shipping crate to unpack the instrument You should receive one box containing the thermal cycler and the accessories 2 Remove the packing material then inspect the instrument for shipping damage 3 Use compressed air to clear out each well of the sample block to remove pa...

Page 2: ...t Mobile Application on your mobile device a For iPad or iPhone devices download the application from the iTunes music store by searching for Instrument Connect by Thermo Fisher Scientific b For Android devices download the application from Google Play by searching for Instrument Connect by Thermo Fisher Scientific 2 Launch the Instrument Connect Mobile Application and log in using your Connect lo...

Page 3: ...he VeritiPro Thermal Cycler User Guide Pub No MAN0019157 How to use the MicroAmp 96 well Tray and Retainer IMPORTANT Use the MicroAmp 96 well Tray and Retainer to prevent crushing of tubes in the thermal cycler When small number of tubes 1 2 tube strips or 8 16 tubes are placed in the sample block without support they can be crushed or deformed when the lid of the thermal cycler is closed over the...

Page 4: ...mp 8 Tube Strip with Attached Caps 0 2 mL 2 MicroAmp 96 Well Tray 3 MicroAmp Splash Free 96 Well Base Seal tube strips with cap strips IMPORTANT Apply significant downward pressure on the sealing tool in all steps to form a complete seal on top of the tubes 1 Align and place the cap strips on the tubes 2 Seal the cap strips using the rocking capping tool a Slip your fingers through the handle with...

Page 5: ...t for connecting to a network USB v2 0 port for connecting to an external network drive jump drive or other USB storage device USB v2 0 port for connecting to USB enabled Wi Fi Card Life Technologies Holdings Pte Ltd Block 33 Marsiling Industrial Estate Road 3 07 06 Singapore 739256 For descriptions of symbols on product labels or product documents go to thermofisher com symbols definition The inf...

Reviews: