background image

VeritiPro

 Thermal Cycler Installation and Operation

Catalog Number  

A47394

Doc. Part No.  

 100093319    

Pub. No.  

 MAN0019367   

Rev.  

 A.0

Note: 

For safety and biohazard guidelines, see the “Safety”

appendix in the 

VeritiPro

 Thermal Cycler User Guide

(Pub. No. MAN0019157). Read the Safety Data Sheets (SDSs) and

follow the handling instructions. Wear appropriate protective

eyewear, clothing, and gloves.

Product description

This document summarizes procedures for installing and using the

VeritiPro

 Thermal Cycler with 6-zone 96-well 0.2 mL block and

VeriFlex

 "better than gradient

" technology. For detailed

instructions, see the 

VeritiPro

 Thermal Cycler User Guide

 (Pub. No.

 MAN0019157).

Fig. 1  VeritiPro

 Thermal Cycler

Required materials not provided

• Can of compressed air (For use in cleaning wells of sample

block).

(Optional)

 Electrical protective devices.

Note: 

The use of one or more of the following electrical

protective devices is recommended.

·

Power line regulator (100–240 V)

·

Surge protector/line conditioner (10-kVA)

·

Uninterruptible power supply (1.5-kVA)

Set up the VeritiPro

 Thermal Cycler

IMPORTANT! 

Save the packing materials and box in case you

need to ship the instrument in for service.

1.

Open the shipping crate to unpack the instrument. You should

receive one box containing the thermal cycler and the

accessories.

2.

Remove the packing material, then inspect the instrument for

shipping damage.

3.

Use compressed air to clear out each well of the sample block

to remove particles that may have collected inside during

shipping.

4.

Move the instrument to an installation site that meets the

spatial and weight requirements for the thermal cycler (see 

“Technical specifications” on page 5).

5.

Connect the instrument to a power outlet.

a.

Connect the power cord to the instrument.

1

1

Power cord port

b.

Install any desired electrical protective devices.

c.

Connect the power cord to a wall plug.

6.

Turn the power switch on, then wait for the instrument to start

up. Proceed with the installation after the touchscreen displays

the Home screen, indicating that the instrument is active.
When you power on the instrument, the instrument will take

under a minute to start up.

1

1

Power switch

The Home screen will display following successful installation.

QUICK REFERENCE

For Research Use Only. Not for use in diagnostic procedures.

Summary of Contents for Applied Biosystems VeritiPro

Page 1: ...ro Thermal Cycler IMPORTANT Save the packing materials and box in case you need to ship the instrument in for service 1 Open the shipping crate to unpack the instrument You should receive one box containing the thermal cycler and the accessories 2 Remove the packing material then inspect the instrument for shipping damage 3 Use compressed air to clear out each well of the sample block to remove pa...

Page 2: ...t Mobile Application on your mobile device a For iPad or iPhone devices download the application from the iTunes music store by searching for Instrument Connect by Thermo Fisher Scientific b For Android devices download the application from Google Play by searching for Instrument Connect by Thermo Fisher Scientific 2 Launch the Instrument Connect Mobile Application and log in using your Connect lo...

Page 3: ...he VeritiPro Thermal Cycler User Guide Pub No MAN0019157 How to use the MicroAmp 96 well Tray and Retainer IMPORTANT Use the MicroAmp 96 well Tray and Retainer to prevent crushing of tubes in the thermal cycler When small number of tubes 1 2 tube strips or 8 16 tubes are placed in the sample block without support they can be crushed or deformed when the lid of the thermal cycler is closed over the...

Page 4: ...mp 8 Tube Strip with Attached Caps 0 2 mL 2 MicroAmp 96 Well Tray 3 MicroAmp Splash Free 96 Well Base Seal tube strips with cap strips IMPORTANT Apply significant downward pressure on the sealing tool in all steps to form a complete seal on top of the tubes 1 Align and place the cap strips on the tubes 2 Seal the cap strips using the rocking capping tool a Slip your fingers through the handle with...

Page 5: ...t for connecting to a network USB v2 0 port for connecting to an external network drive jump drive or other USB storage device USB v2 0 port for connecting to USB enabled Wi Fi Card Life Technologies Holdings Pte Ltd Block 33 Marsiling Industrial Estate Road 3 07 06 Singapore 739256 For descriptions of symbols on product labels or product documents go to thermofisher com symbols definition The inf...

Reviews: