A system dye calibration requires ~30 minutes to complete.
Prepare the dye set calibration standards and plate as described in the product
information sheet for the dye set.
In the home screen:
1.
Touch Settings
4
Maintenance and Service
4
Calibration
4
Dye Calibration.
2.
Touch the injection group for the dye set in the plate, then touch Dye set.
Perform a system
dye calibration
Appendix C
Dye calibration and install standard checks
Calibrate dyes
C
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide