Calibration description and purpose
Pass criteria
Normalization
(TaqMan
™
Array Card block only)
• The software extracts a spectral profile for each normalization standard.
• The software uses the normalization standards to obtain a correction
factor for the specific dyes, for example, FAM
™
dye or VIC
™
dye.
• The correction factor is used to generate comparable component signal
levels across instruments.
Normalization standards have
sufficient signals to generate a
correction factor.
View the calibration status and set reminders
View the calibration status on the instrument
The
Status and Reminders
screen also allows you to perform the following functions:
• Transfer the calibration results (see
• Transfer the calibration data files (see
)
• View the calibration images (see
1.
In the home screen, tap
(Settings)
4
Maintenance and service
4
Calibration
4
Status and
reminders
.
The
Status and Reminders
screen is displayed. It shows the following information:
• Type of calibration
Note:
·
For dye calibration, the type of calibration is listed as the dye that was calibrated, for
example,
FAM
.
·
Normalization calibration is required for a TaqMan
™
Array Card block. The type of calibration
is listed as
FAM/ROX
or
VIC/ROX
.
• Status of the calibration
• Last calibration
• Date that the calibration expires
• If a reminder is set
2.
Tap a calibration row, then tap
View
to view the history of that specific calibration type.
• Date performed
• Result of the calibration
• Whether the calibration is current
• Operator
3.
Tap
Cancel
to return to the
Status and Reminders
screen.