Filters
Wavelength
(nm)
1
2
3
4
5
x1-m1 x2-m2 x3-m3 x4-m4 x5-m5 x6-m6
Emission
Spectra
1
x1-m1—FAM
™
dye, SYBR GREEN
™
dye
2
x2-m2—VIC
™
dye
3
x3-m3—ABY
™
dye, NED
™
dye, Cy
®
3, TAMRA
™
dye
4
x4-m4—JUN
™
dye, ROX
™
dye, Texas Red
™
dye
5
x5-m5—Cy
®
5, MUSTANG PURPLE
™
dye
Custom dyes
The instrument can run assays designed with custom dyes. Custom dyes include the following options:
• Dyes that are not supplied by Thermo Fisher Scientific.
• Dyes or formulations of dyes that are not system dyes for the instrument.
To use a custom dye on the instrument, review the following requirements.
• Calibrate the instrument for the custom dye (see “Calibrate custom dyes”
).
• Ensure that the custom dye excites between 455–672 nm and emits between 505–723 nm.
• Select a custom dye that does not overlap with other dyes used in the run (see “System dyes”
• The custom dye for use in calibration must be attached to the 5' end of a short DNA
oligonucleotide consisting of the first two bases of the probe sequence without a quencher at
the 3' end.
Chapter 1
Product information
Instrument hardware description
1
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide
17