Regulatory and Safety Information
To power off the instrument:
1.
Close any software applications that are open and then log off.
2.
Shut down the computer.
3.
Move the printer power switch to the off position.
4.
Move the instrument power switch to the off position. Leave the power
supply connected unless you will be servicing the instrument.
Fuses
Two fuses protect the power supply from excessive current. Use
replacement fuses which meet the following specifications:
Warning
Avoid shock hazard and overheating. Always use an exact replacement for
the fuse.
Fuse characteristic
Specification
Current (max)
2 A
Voltage 250
V
Size
6.3 mm x 32 mm
0.25 in x 1.25 in
Action Fast
acting
Warning
To avoid the risk of electrical shock, always turn off the instrument and
disconnect it from the wall outlet or power strip before changing fuses.
▲
Thermo Fisher Scientific
Antaris II and Antaris MX Site and Safety Information
17