Antaris EX-IP Specifications
Floor mounting
Optionally, a system may be mounted to the floor. The floor must be
capable of supporting two times the weight of the instrument and floor
stand.
In this configuration, the wall-mount flanges may be removed.
The instrument must be accurately secured within 2 degrees of vertical.
The floor used for mounting must be within 2 degrees of level for proper
operation.
All 4 bolts are required to safely secure the instrument to the stand. For
anchoring the enclosure, we recommend using high quality steel 10 mm
(3/8 in) bolts that are capable of withstanding the following forces:
Bolt characteristic
Specification
Shear force
250 N (56 lb)
Tensile force
300 N (68 lb)
Pull-out force
30 N (68 lb)