Antaris EX-Z2 Specifications
Shelf or
table top mounting
Optionally, a system may be mounted to a table top or shelf. The surface
must be capable of supporting two times the weight of the instrument. See
the section entitled “Lifting or moving” in the “Safety” chapter earlier in
this manual for lifting/moving precautions.
The wall mount flanges may be removed.
The equipment must be on a surface that is within 2 degrees of level for
proper operation.
For anchoring the enclosure, we recommend using high quality steel
10 mm (3/8 in) bolts that are capable of withstanding the following forces:
Force
Specification
Shear force
250 N (56 lb)
Tensile force
300 N (68 lb)
Pull-out force
30 N (68 lb)
94
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific