Change EVOS
™
light cubes
Install or change an EVOS
™
light cube
The Countess
™
3 FL Automated Cell Counter can accommodate up to two EVOS
™
light cubes. Each
user-interchangable, auto-configured EVOS
™
light cube contains an LED, collimating optics, and filters
for fluorescence applications. EVOS
™
light cubes do not come standard with the device and must be
purchased separately (“EVOS
). To install or change a light cube:
1.
Press
Settings
on the
Home
page.
2.
Press
Instrument Settings
.
3.
Press
Change light cube
. The instrument positions the light cube tray to enable light cube
installation.
4.
When prompted, power off the Countess
™
3 FL Automated Cell Counter using the power switch on
the back of the instrument (“Exterior instrument parts”
).
5.
Unplug the power cord from the Countess
™
3 FL Automated Cell Counter.
6.
Unlatch the back panel with the two captive ¼-turn fasteners (indicated by black arrows below)
that secure the back panel on the rear of the Countess
™
3 FL Automated Cell Counter and remove
the back panel.
10
Countess
™
3 FL Automated Cell Counter User Guide
51