Alpha pH 1000

Controller/Transmitter

pH / ORP

Summary of Contents for Alpha pH 1000

Page 1: ...Alpha pH 1000 Controller Transmitter pH ORP...

Page 2: ...k ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Star Navigator 21 Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are trademarks of Ther...

Page 3: ...applications of the Alpha pH1000 pH ORP process controller transmitter If you have doubts in the use of the instrument please do not hesitate to contact the nearest Authorized Distributor Thermo Scien...

Page 4: ...tion of the instrument Before powering up the unit make sure that power source is connected to it as specified on the top label Failure to do so may result in permanent damage of the unit The unit has...

Page 5: ...ion 13 7 ADVANCED SET UP MODE _________________________________________ 14 7 1 Electrode Offset OFS sub function 14 7 2 Setting temperature Set o C sub function 14 7 3 Control Relay A Control Relay B...

Page 6: ...rical symmetrical input for pH ORP operation Three control modes limit proportional pulse length or proportional pulse frequency Large dual display LCD for easy reading with clear multiple annunciator...

Page 7: ...matching pin PMP a final control element such as pump or valve a recorder an appropriate pH or ORP measurement cable PH Controller Flow Assembly Measurement Cable Process Assembly with Electrode Powe...

Page 8: ...ore proceeding Signal Output Input Hold Alarm Relay B Relay A Power Mains 3 pH mA PE AC N L 1 2 5 4 3 6 7 8 9 10 11 12 13 indicated contact positions are for currentless con Pt 100 Potential Matching...

Page 9: ...BNC 0 PA PE N L S 22 21 1 2 3 20 19 18 Potential Matching Pin PMP s Symmetrical high impedance connections Te 0 PA PE N L S 22 21 1 2 3 20 4 Asymmetrical connections Measuring Cable pH Combination Ele...

Page 10: ...ction 8 High set relay common 17 4 20 mA for ve connection 9 High set relay working position NO Connections for the 5 way screw terminals 18 Pt100 Pt1000 lead 1 terminal 19 Pt100 Pt1000 sense lead ter...

Page 11: ...relay operation mode 4 1 2 Display The LCD display features two numerical displays that show status messages and measured values for easy quick reference The display provides short text information f...

Page 12: ...1 How to view operating parameters without access to change them a Press the Enter key The display will prompt the user to enter a security code S Cd Leave the security code at 000 do not enter a sec...

Page 13: ...Controller type N O N C Contact Current output 0 20 4 0 mA Cycle time bzw Max Frequency pH ORP at 0 4 mA Proportional range XP pH ORP Symm Asymmetrical Alarm delay Alr Alarm contact cont wiper cont R...

Page 14: ...ey without adjusting the security code leave code at 000 The upper display shows electrode slope The lower display reading shows the pH reading at 0 mV Note If security code is changed to a value othe...

Page 15: ...scroll the display to Setup security code 22 3 Press the ENTER key 4 If configured for pH control the upper display reads OFS 5 If configured for ORP control the upper display reads SP1 NOTE Pressing...

Page 16: ...s used for the first point calibration It would either be 7 00 or 6 86 pH pH pH 3 Press the or key to scroll the lower display to the buffer value that matches your buffer solution 4 Make sure the ele...

Page 17: ...etting the calibration temperature and the process temperature see section 7 2 3 6 2 ORP mV Calibration This mode allows one point calibration 1 Enter Calibration mode While in the Measurement mode pr...

Page 18: ...e value of 80 4 Press the ENTER key to confirm the value The electrode indicator and CAL indicator flash The unit automatically adjusts its reading to match the solution value 5 The display will now s...

Page 19: ...ator appears at the top of the display The lower display shows the current measured pH value The upper display shows the current offset value 6 Press the or key until the pH value coincides with the c...

Page 20: ...anual temperature compensation two different temperatures process and calibration can be input independently This allows calibration at a temperature other than the process temperature Example setting...

Page 21: ...set point to activate controller when the pH ORP value goes below the low set point select a high set point to activate controller when the value overshoots the high set point Using both SP1 and SP2...

Page 22: ...on delay time and the lower display shows On d 0 SETUP HOLD 3 Press the or key to enter on delay time for Set point 1 Set point 2 The controller will delay activation for the number of seconds 0 to 20...

Page 23: ...se length proportional control for precise control of proportioning valves Use control off to operate controller as a monitor only or to keep relays from switching 1 Follow directions in 7 4 1 to ente...

Page 24: ...ing the and keys simultaneously escape 7 4 5 Maximum Pulse Length tPL or Maximum Frequency FPF Note If the controller type oFF or L Ct is set the parameters listed in 7 4 4 and 7 4 5 are blanked out T...

Page 25: ...mode ORP parameters are blocked out If you are in the ORP mode pH parameters are blocked out 1 Follow directions in 7 5 1 to enter Current Output mode 2 Press the ENTER key Scroll with the or keys un...

Page 26: ...ORP or ORP mV measurement This parameter group lets you select pH mV or readings and input type symmetrical or asymmetrical Symmetrical mode is used in conjunction with a potential matching pin in an...

Page 27: ...splay shows a numerical value in seconds and the lower display shows AL d 3 Press the or key to select how long of an alarm delay 0 to 2000 seconds you want 4 Press the ENTER key to confirm your selec...

Page 28: ...eted 10 TUP LD SE HO 7 6 7 Selecting pH sensor type Note This parameter is blocked out in ORP mode Proceed to 7 6 6 Note After each changeover the factory calibration data for zero point and slope are...

Page 29: ...o factory defaults 4 Press the ENTER key to confirm your selection and to return to Advanced Set up mode or return to Measurement mode by pressing the and keys simultaneously escape Cal TUP LD SE HO S...

Page 30: ...ey lights up Note Pressing ENTER key at a value other than 22 will cause the controller to revert to Measurement mode and the relays will remain in automatic mode 4 Press the RELAY SELECTION key to se...

Page 31: ...acts Switching voltage current power max 250 VAC max 3A max 600 VA Alarm Functions Function switchable Latching pulse Wash cycle 0 1 to 199 9 hrs Wash duration 1 to 1999 s Pickup delay 0 to 2000 s Swi...

Page 32: ...hanical Specifications Dimensions control panel housing L x H x W 175 x 96 x 96 mm Weights control panel housing max 0 7 kg Material ABS with polycarbonate front housing Insulation Front Housing IP 54...

Page 33: ...SOHF05B pH Combination Electrode with PMP w o ATC ECARGTSO05B pH Combination Electrode with high temperature resistance 110 oC and high pressure resistance 9 bar w o ATC PMP ECARHTTSO05B pH Combinatio...

Page 34: ...nit rubber gasket 1 unit 11 3 Return of Goods Authorisation must be obtained from our Customer Service Dept to issue a RMA Return of Material Authorisation number before returning items for any reason...

Page 35: ...the power supply Jumper Positions Internal to the controller JP 1 Selects the input voltage 220 VAC JP 2 Selects the input voltage 110 VAC JP 3 Solder bridge selects between Pt100 and Pt1000 Fuse Note...

Page 36: ...32 10 18 15 0 99 4 00 6 90 7 04 9 11 9 27 10 12 20 1 00 4 00 6 88 7 02 9 05 9 22 10 06 25 1 01 4 01 6 86 7 00 9 00 9 18 10 01 30 1 01 4 01 6 85 6 99 8 95 9 14 9 97 35 1 01 4 02 6 84 6 98 8 91 9 10 9 9...

Page 37: ...FORWARD DIRECTION REVERSE DIRECTION SP1 RELAY ON SP2 SP1 Set to LO SP2 Set to HI The controller relay activates when the set point is reached In the reverse direction it does not de activate when the...

Page 38: ...ction Control characteristic of P Controller as proportional controller Xp 0 0 50 100 Yh Xw SP 1 MIN function Xp 0 Xw SP 2 MAX function Control characteristic of P Controllers as limit value switch Ch...

Page 39: ...r tON Xp exceeded tON T relay remains picked up Relay ON tON tOFF Pulse Length T OFF Time s Controller signal of Pulse Frequency Controllers The output relay of the pulse frequency controller is clock...

Page 40: ......

Page 41: ......

Page 42: ...Fax 978 232 6031 Europe P O Box 254 3860 AG Nijkerk Wallerstraat 125K 3862 CN Nijkerk Netherlands Tel 31 033 2463887 Fax 31 033 2460832 Asia Paci c Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949...