AI 3000/AS 3000 II Control
Control Through the TRACE GC Ultra
62
AI 3000/AS 3000 Sampling System - Operating Manual
Thermo Scientific
Control Through the TRACE GC Ultra
This paragraph describes how to program and control the AI 3000/AS 3000 II through the
TRACE GC Ultra keypad.
Groups of samples can be automatically analyzed under different analytical conditions by
programming a
sequence
of samples. The sequence includes a set of information for a group of
samples and their position on the sample tray. The sequence also specifies the method to be
used to run each group of samples.
Configuring the AI 3000/AS 3000 II
When the GC is switched on, the presence of the autoinjector/sampler and its configuration is
automatically acknowledged.
The type of configuration is displayed by pressing
CONFIG
and selecting
AUTOSAMPLER
.
The following non editable message is displayed according to the AI 3000/AS 3000 II
installed:
for the AS 3000.
or
for the AI 3000.
Note
It is not possible to control the 0.5
μ
L syringe through the GC keypad but only
through the data system.
Note
In an analytical method all the AI 3000/AS 3000 II functions can be programmed.
To develop an analytical method and activate a sample sequence, please refer to the
TRACE GC Ultra Operating Manual
.
CONFIG AUTOSAMPLER
Sample tray
105
CONFIG AUTOSAMPLER
Sample tray
8
Note
Should the AI 3000/AS 3000 II not be installed or correctly connected, the TRACE
GC Ultra will display the message NO AUTOSAMPLER INSTALLED.