Installation
Setup
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
2-9
dry run by attaching the heated sample tube on top of the 5028
i
base. If
the heated tube protrudes through the roof flange, either the instrument
should be lowered or an elongated PVC sleeve and cap system should
cover the heater while maintaining a waterproof seal. However, for the
remainder of this description, it is assumed the heater will fit within the
ceiling height of the air monitoring station.
Based on the vertical height requirements, use additional 5/8-inch OD
stainless tubing and connectors to make up the difference between the
desired inlet height and the top of the heater.
Additional routing is necessary for the ambient T/RH sensor assembly.
Ambient Inlet
(PM
10
or PM
2.5
)
Ambient T/RH
Sensor Assembly
Tripod Adapter
T/RH Sensor Cable
Tripod
Roof Flange
5/8” SS Tubing
5/8” Union
w/Teflon Ferrule
3-Feet Universal
Heater
Figure 2–4.
Model 5028
i
Vertical View
Sample tube extensions are used for added height. Two sample tube
extensions for the Model 5028i separate vertically by minimum
requirement of 18 Inches entrance-to-entrance. Usually 40 Inches from
inlet top of one Channel to another inlet top is recommended.
Sample Tube Extension
Lengths
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......