System Description
Electronics
8-6
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
All electronics operate from a universal switching supply, which is
capable of auto-sensing the input voltage and working over the all
specified ranges.
External pumps and heaters all operate on 220 VAC. An optional
transformer is required if operating on the 110VAC.
An on/off switch controls all power to the analyzer and is accessible on
the front panel.
The motherboard contains the main processor, power supplies, and a
sub-processor, and serves as the communication hub for the instrument.
The motherboard receives operator inputs from the front panel function
keys and/or over I/O connections on the rear panel. The motherboard
sends commands to the other boards to control the functions of the
instrument and to collect measurement and diagnostic information.
The motherboard outputs instrument status and measurement data to
the graphics display and to the rear-panel I/O. The motherboard also
contains I/O circuitry and the associated connector to monitor external
digital status lines and to output analog voltages that represent the
measurement data. Connectors located on the motherboard include:
External connectors include:
l
External Accessory
l
RS-232/485 Communications (two connectors)
l
Ethernet Communications
l
I/O connector with Power Fail Relay, 16 Digital Inputs, and 6
Analog Voltage Outputs.
Internal connectors include:
l
Function Key Panel and Display
l
Measurement Interface Board
l
I/O Expansion Board
l
Digital Output Board
l
AC Distribution
The measurement interface board serves as a central connection area
for all measurement electronics in the instrument. It contains power
supplies and interface circuitry for sensors and control devices in the
measurement system. It sends status data to the motherboard and
receives control signals from the motherboard.
Connectors located on the measurement interface board include:
l
Data communication with the motherboard
Electronics
Motherboard
Measurement Interface
Board
Measurement Interface Board
Connectors
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......