Operation
Service Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-81
The Alpha Threshold screen allows the user to view and set the alpha
threshold manually.
l
In the Main Menu, choose Service > Detector Calibration >
>
Detector Calibration or Detector Calibration B
> Manual >
Alpha
Threshold
.
ALPHA THRESHOLD:
CURRENTLY: 1500
SET TO: 1501 ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see “
in this chapter.
Note
This adjustment should only be performed by an instrument
service technician.
s
l
In the Main Menu, choose Service >
Analog Out Calibration
.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
Alpha Threshold
Analog Output Calibration
Summary of Contents for 5028i
Page 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Page 2: ......
Page 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Page 6: ...WEEE Compliance Thermo Fisher Scientific...
Page 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Page 36: ......
Page 142: ......
Page 154: ......
Page 220: ......
Page 232: ......
Page 328: ......
Page 329: ......