3-21
Reset User Calibration Default
The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
•
In the Main Menu, choose Service > Calibration Factors >
Reset User
Cal Defaults
.
•
Press
to warn user and enable restore with
.
•
Use
to reset the calibration pressure reference value when pressed
after .
Calibration Menu
The Calibration menu is used to calibrate zero and span. The zero/span
check is displayed only if the sample/cal valve option is installed. The
screens below show the calibration menu in single mode and dual/auto
range modes.
The calibration menu is similar for the single, dual, and auto range mode,
however, the dual and auto range modes have two O
3
span factors (high and
low). This allows each range to be calibrated separately. This is necessary if
the two ranges used are not close to one another. For example, a low O
3
range of 50 ppb and a high O
3
range of 20,000 ppb. For more information
about calibration, see Chapter 4, “Calibration”.
O3 COEFFICIENT:
O3: 600
SET COEF TO: 1.000
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE DEFAULT CAL:
RESTORE
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...