C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-15
set alarm pressure min
value
set alarm pressure max
value
These commands set the pressure alarm minimum and maximum values to
value,
where
value
is a floating-point number representing pressure alarm
limits in millimeters of mercury. The example that follows sets the pressure
alarm maximum value to 1000 mmHg.
Send:
set alarm pressure max 1000
Receive:
set alarm pressure max 1000 ok
alarm sample flow a min
alarm sample flow a max
alarm sample flow b min
alarm sample flow b max
These commands report the cell a and b sample flow alarm minimum and
maximum value current settings. The example that follows reports that the
sample flow alarm minimum value is 0.400 LPM.
Send:
alarm sample flow min
Receive:
alarm sample flow min 0.4 l/min
set alarm sample flow a min
value
set alarm sample flow b max
value
set alarm sample flow a min
value
set alarm sample flow b max
value
These commands set the cell a or b sample flow alarm minimum and
maximum values to
value,
where
value
is a floating-point number
representing sample flow alarm limits in liters per minute. The example that
follows sets the sample flow alarm maximum value to 1.400 LPM.
Send:
set alarm sample flow max 1.4
Receive:
set alarm sample flow max 1.4 ok
alarm trig conc o3
This command reports the O
3
concentration alarm trigger action for
minimum alarm, current setting, to either floor or ceiling. The example that
follows shows the O
3
concentration minimum alarm trigger to ceiling,
according to
Send:
alarm trig conc 03
Receive:
alarm trig conc 03 1
set alarm trig conc o3
value
These commands set the O
3
concentration alarm minimum
value,
where
value
is set to either floor or ceiling, according to
that follows sets the O
3
concentration minimum alarm trigger to ceiling.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...