Servicing
Analog Input Calibration
Thermo Fisher Scientific
Model 49
i
Instruction Manual
7-15
2. From the Main Menu, press
to scroll to Service, press
>
to scroll to Analog Out Calibration
,
and press
.
The Analog Output Cal screen displays
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
3. At the Analog Output Cal screen, press
to scroll to the voltage
channel or current channel corresponding to the rear panel terminal
pins where the meter is connected, then press
.
4. With the cursor at Calibrate Zero, press
.
The Analog Output Cal line displays Zero
Note
When calibrating the analog output, always calibrate zero first and
then calibrate full scale.
5. Use
until the meter reads the value shown in the Set
Output To line, then press
to save the value.
6. Press
to return to the previous screen.
7. Press
to select Calibrate Full Scale.
8. Use
until the meter reads the value shown in the Set
Output To line, then press
to save the value.
Analog Input
Calibration
Use the following procedures to calibrate the analog inputs after replacing
the optional I/O expansion board. These procedures include selecting
analog input channels, calibrating them to zero volts, and then calibrating
them to full scale using a known voltage source.
Calibrating the Input
Channels to Zero Volts
Use the following procedure to calibrate the input channels to zero volts.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...