Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-83
Service Menu
The Service menu appears only when the instrument is in the service mode.
To put the instrument into service mode:
•
In the Main Menu, choose Instrument Controls >
Service Mode
.
Advanced diagnostic functions are included in the service mode.
Meaningful data should not be collected when the instrument is in the
service mode.
•
In the Main Menu, choose
Service
.
Pressure Check
The Pressure Check menu is used to manually control the flow of reference
or sample gas through Cell B. This enables the pressure reading of Cell B,
with either sample or reference gas, to be determined. Pump pressure is used
to test the pump. Selecting any of these menu items will disturb the analog
outputs.
MIN TRIG(CEILING/FLOOR):
ACTUAL TRIGGER: CEILING
SET TRIGGER TO: FLOOR ?
TOGGLE AND SAVE VALUE
RANGE AVG DIAGS ALARM
SERVICE:
>PRESSURE CHECK
LAMP SETTING
DETECTOR CALIBRATION
INTENSITY CHECK
RANGE MODE SELECT
PRESSURE CALIBRATION
FLOW A CALIBRATION
ª
RANGE AVG DIAGS ALARM
FLOW B CALIBRATION
TEMPERATURE CALIBRATION
ANALOG OUT CALIBRATION
ANALOG INPUT CALIBRATION
DILUTION RATIO
DISPLAY PIXEL TEST
RESTORE USER DEFAULTS
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...