Servicing
Optional DMC Permeation Oven Solenoid Valve Replacement
7-42
43iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the DMC permeation oven solenoid
valve option.
Equipment required:
Phillips drive, #1
CAUTION
Allow the oven to cool down prior to servicing.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
2.
Unplug perm valve J7 from step pol board 2.
Figure 7–48.
Removing and Replacing the Solenoid Valve pt 1
3.
Loosen two 4-40 screws and slide optional perm valve out.
Optional DMC
Permeation Oven
Solenoid Valve
Replacement