![Thermo Scientific 42iQD Instruction Manual Download Page 245](http://html.mh-extra.com/html/thermo-scientific/42iqd/42iqd_instruction-manual_1105143245.webp)
System Description
Peripherals Support System
Thermo Scientific
42iQD Instruction Manual
8-7
The peripherals support system operates these additional devices that are
needed, but do not require special feedback control or processing. These
components are connected to a Peripherals Support Board (PSB).
The chassis fan provides air cooling of the active electronic components.
The Step POL board provides high/low outputs for continuous operation
or on/off states. The Step POL board contains the basic circuitry to provide
a programmable load to passive devices, either continuously, or on user or
automated command. In the iQ Series instruments, the pump, solenoids,
etc., are controlled off of the Step POL board from commands generated
via the PSB.
The vacuum pump draws the reacted gases out of the reaction chamber.
Optional solenoid valves for switching between sample, zero, and span
gases, and other optional components.
The flow pressure DMC performs the pressure measurements that assure
proper flow regulation, and also for the sample pressure within the reaction
chamber. It is coupled with a standard restricting capillary for flow when
the downstream pump is operating: Upstream pressure is ambient, while
downstream pressure allows determination of flow. A single PCBA with
microprocessor provides the active controls for the pressure measurements,
performs flow determination, and generates registers that interact with the
higher level system controls.
Peripherals
Support System
Fan
Step POL Board
Pump
Solenoid Valve
Panel
Flow/Pressure
DMC