Servicing
Optional Manifold Replacement
7-46
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to replace the manifold.
Equipment required:
Hex wrench, 9/16
Hex drive, 9/64
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
2.
Unplug three electrical connections (J5, J6, and J8) from the Step POL
board 1.
Figure 7–50.
Replacing the Manifold pt 1
3.
Unfasten three nuts. Remove the nuts, front and back ferrules as shown
from span, zero in, sample back panel (
Optional
Manifold
Replacement