Servicing
DMC PMT Cooler and Reaction Chamber Replacement
7-30
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to clean or remove the reaction chamber.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove cooler/reaction chamber DMC as described in “
” in this chapter.
3.
Unplug heater J8 from board.
Figure 7–34.
Photo Reaction Chamber Cleaning, Electrical
4.
Disconnect plumbing as shown.
5.
Using a 9/64-inch hex drive, unfasten three #8-32 socket cap head
screws.
Reaction Chamber
Cleaning and/or
Removal