Intended Use
The Refrigerators/Freezers described in this manual are for
professional use only. These products are intended for use in
research for the storage of samples or inventory in the
following temperature ranges:
+1°C to +12°C
3
W
ARNING
Only Explosion Proof Units or
Flammable Material Storage Units
are to be used for the storage of
flammable inventory/samples.
These are not considered medical devices and have therefore not
been registered with a medical device regulatory body (e.g. FDA):
that is, it has not been evaluated for the storage of samples for
diagnostic use or for samples to be re-introduced to the body.
Corrosive Materials Requiring Refrigerated or Frozen Storage
Refrigerators
Freezers
For 20EFEETSV
-
12°C to -20°C
-
18°C to -28°C
• Only use models rated as Corrosion Resistant
.
• Corrosion Resistant does not mean Corrosion Proof – Care in
storage
is still required
.
• Store only corrosive reagents/samples which truly need
reduced temperature storage.
• Flammable corrosive materials require Flammable Material Storage
or
Explosion Proof models labeled for the storage of corrosives
.
• Containers must be wiped clean of moisture and chemical
residue
before being introduced into the unit and upon return
from use.
• Containers must be sealed with either vinyl tape or Parafilm ®
.
• Reagents/samples which release HX (X= F, Cl, Br, I) on contact
with
moisture (e.g. Acyl halides, Organosilyl halides etc.) are
particularly
damaging to metals.
• Volatile amines will react with HX depositing salts which will
lead to corrosion of metal surfaces.
• Volatile organic acids can exacerbate metal surfaces
already
compromised, be sure these are securely sealed.
• Refrigerated compartments are cool or cold areas but they are not
to
dry seal your reagents and samples.
• Bleach solutions release chlorine gas which can react with
other volatiles in the cooling chamber or directly attacks metal
surfaces.
• Periodically clean the interiors, clean up spills or leaking containers
• Failure to take precautionary actions may lead to damage not
covered
by warranty claims
.