Troubleshooting
Troubleshooting Guides
Thermo Fisher Scientific
Model 146
i
Instruction Manual
6-5
Contaminated cell
Clean cell.
Sticky solenoid valve
Replace with known good
solenoid valve.
Photometer does not
calibrate properly
Leak
Perform leak test (see
“Servicing” chapter.
Bench thermistor defective
Replace bench thermistor.
Pressure transducer out of
calibration
Recalibrate pressure
transducer.
Dirty system
Clean cells and flow
components.
Defective solenoid
Replace solenoid.
Lamp intensity unstable
Replace detector or
photometer board.
Slow response
Averaging time
Verify averaging time is set
properly.
Leak
Perform leak test (see
“Servicing” chapter).
Contaminated optical bench
Clean bench and then
condition system overnight.
Table 6-2.
Troubleshooting - Alarm Messages
Alarm Message
Possible Cause
Action
Alarm - O
3
Lamp Temp
Defective ozonator board
Replace ozonator board.
Defective ozonator lamp heater
Replace ozonator lamp
heater.
Alarm - Lamp Temp
Defective photometer board
Replace photometer board.
Defective photometer lamp heater
Replace photometer lamp
heater.
Alarm - Bench Temp
Defective fan
Check for 24 VDC on fan
connector and replace fan if
24 VDC is applied to fan.
Check fan filter
Clean or replace foam filter
(see “Preventive
Maintenance” chapter).
Table 6-1.
Troubleshooting - General Guide, continued
Malfunction
Possible Cause
Action
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...