Dionex PDA Operator’s Manual
66
Doc. 065378-01 1/12
To troubleshoot:
This message appears if you attempt to issue a command or select a parameter
while a diagnostic routine is running. Wait until the diagnostic routine is
completed before making any changes.
To troubleshoot:
This message appears if there is a communication failure between the DSP
(digital signal processor) and the CPU. Turn off the Dionex PDA power for
30 seconds, and then turn it on again. If the error message appears again,
contact Thermo Fisher Scientific for assistance.
A backup paddle alignment routine ensures that the Dionex PDA remains
functional even if the sensor cannot detect the filter wheel position. Thus, if
this is the only error reported, you may continue operation and defer
troubleshooting.
To troubleshoot:
1. Turn off the Dionex PDA power for 30 seconds, and then turn it on again.
2. If the error message appears again, check the deuterium lamp intensity:
a. Open the Chromeleon 7 Command window (see
) or the
Chromeleon 6.8 Commands dialog box (see
).
Diagnostic is running.
DSP communication failure.
Filter wheel sensor could not detect blocked position.
Filter wheel sensor could not detect open position.
Summary of Contents for 074114
Page 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Page 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Page 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Page 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Page 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Page 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Page 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...