Figure 2-2. Dionex PDA Status Bar
LED Label
LED Status
Function
CONNECTED
Lighted
The Dionex PDA is connected to a Chromeleon 7
instrument or a Chromeleon 6.8 timebase.
Off
The Dionex PDA is not connected to a
Chromeleon 7 instrument or a Chromeleon 6.8
timebase.
ALARM
Lighted
A problem has occurred (a leak, for example).
Check the Chromeleon Audit Trail for the cause.
Off
No problems have been detected.
DEUTERIUM – UV
Lighted
The deuterium lamp is on.
Flashing
The LED flashes once every second while the
deuterium lamp is firing.
The LED flashes every 0.5 second if a command
is issued to turn on the lamp, but the lamp fails to
turn on (see
). If a subsequent attempt
to turn on the lamp is successful, the LED will
automatically stop flashing.
Off
The deuterium lamp is off.
TUNGSTEN – VIS
Lighted
The tungsten lamp is on.
Flashing
The LED flashes once every second while the
tungsten lamp is firing.
The LED flashes every 0.5 second if a command
was issued to turn on the lamp, but the lamp failed
to turn on (see
). If a subsequent
attempt to turn on the lamp is successful, the LED
will automatically stop flashing.
Off
The tungsten lamp is off.
POWER
Lighted
The Dionex PDA power is on.
Off
The Dionex PDA power is off.
CONNECTED
ALARM
DEUTERIUM - UV
POWER
TUNGSTEN - VIS
Summary of Contents for 074114
Page 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Page 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Page 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Page 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Page 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Page 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Page 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...