Service Maintenance
________________________________________ Cleaning the Ion Optics and Mass Analyzer
__________________ LCQ Deca XP Plus Hardware Manual __________________
5-7
Removing the Ion Optics and Mass Analyzer
The ion optics and mass analyzer are mounted on a baffle on the underside of
the top cover plate of the vacuum manifold.
Use the following procedure to remove the ion optics and mass analyzer from
the top cover plate. See Figure 5-2 for the location of the ion optics and mass
analyzer components.
Figure 5-1.
Electrical connections and damping gas line connection to
the top cover plate of the vacuum manifold
Damping
Gas Line
Electron Multiplier
High Voltage Cable
Electrometer Cable
Endcap Electrode
Cables
Octapole/Quadrupole
Cable
Lenses
Cable