Texmate, Inc. Tel. (760) 598-9899 • www.texmate.com
UM-40-ACV manual (UM11)
Page 1
Texmate, Inc. Tel. (760) 598-9899 • www.texmate.com
Page 1
General Features
Compatibility
Specifications
Typical Application Connections
The UM-Series NEMA case style is com
-
plementary to Texmate’s Classic RP-Series.
For economy, each UM model is dedicated
to a specific application. UMs are ideal for
upgrading or replacing the traditional USA
NEMA case panel meters presently in use.
L
N
UM-40AC
SPAN
UM-40AC
SPAN
ZERO
Optional
SI
G
HI
SI
G
LO
SI
G
HI
SI
G
LO
AC Voltage
measurement in Single-phase Systems.
AC Phase to Phase Voltage
measurement
in Multi-phase Systems.
T
S
R
The unique differential input allows safe
phase to phase AC line voltage measurements up to 500V AC.
Maximum display of 500.0V AC.
Maximum display of 500.0V AC.
May be used with 3 Phase 3 Wire
and 3 Phase 4 Wire Systems.
AC Volts
up to
500V AC
AC Volts
up to
500V AC
ZERO
Optional
Differential
Hi Voltage
Protection
2M
2M
4K32
20K
20K
Differential
to Single
Ended
Conversion
Differential
Hi Voltage
Protection
2M
2M
4K32
20K
20K
Differential
to Single
Ended
Conversion
UM-40-ACV
700.0V AC Meter
4 DIGIT with 0.56” LEDs
in a NEMA type 1 Style Case
A utility meter for high resolution AC volts measurement with safe
resistively isolated differential inputs.
The UM-40-ACV is a cost-effective, utility, AC voltage measur
-
ing meter with range of 0 to 700.0V AC. The unique resistively
isolated differential input of this meter allows safe measurement
of phase to phase voltages, making it a effective solution to
most high resolution AC voltage measuring applications.
The standard meter is provided with TB-KIT screw terminal
blocks and insulated quick- disconnects. For the greatest con
-
venience and ease of use, order the optional pre-configured
Push-On screw terminal connectors. (see Push-On Screw
Terminals and Ordering Information)
Input Configuration:
........
Differential input. Inputs resistively isolated
to 1400V from internal ground of meter
by 1.94MΩ, so that phase to phase
measurements up to 700V AC can be
safely made.
Full Scale Ranges:
..........
700V AC full scale
Input Impedance:
.............
4MΩ minimum.
A/D Converter:
.................
16 Bit Dual Slope
Accuracy:
.........................
±(0.05% of r 3 counts)
Temperature Coefficient:
100 ppm/°C (Typical)
Warm Up Time:
................
2 minutes to specified accuracy
Conversion Rate:
.............
3 conversions per second (Typical)
Display:
.............................
4 digit 0.56" Red LED display (standard),
Green or Super Bright Red (optional).
Range 0 to 9999 counts.
Decimal Selection:
..........
Header under face plate, X•X•X•X•
Overrange Indication:
.....
Display flashes “0000”
Power Supply (std):
.........
120/240V AC, 50/60/400Hz. approx 1.5W.
(Optn) V0-DC/ISO
..........
Isolated Switcher. 9 to 36V DC/12 to 24V AC
(Optn) V0-24V
................
Isolated Transformer 24V AC ±10%
Operating Temperature:
..
–10 to 50 °C
Storage Temperature:
.....
–20 to 70 °C.
Relative Humidity:
...........
95% (non-condensing)
Case Dimensions:
...........
Bezel 3.78”Wx1.89”H (96Wx48Hmm)
Depth behind bezel 3.36" (85.3 mm) Plus
0.5 to .9” (12.7 to 22.8mm) depending on
connector used.
Weight:
.............................
NW. 12oz (0.34kg)
15.6oz (0.44kg). when packed.
UM-Series
UM-Series utility meters for switchboard and process indication
UM-35-ACA
....AC amps, Scaled or True RMS, (
1 or 5 Amp
internal shunt), 3.5 digit.
UM-35-ACV
......AC volts, Scaled or True RMS.
199.9V AC
/600V AC header selectable 3.5 digit.
UM-35-DCA
....DC mV ±20mV/±50mV/
±100mV/
±200mV header selectable ranges, 3.5 digit
UM-35-DCV
.....DC Volts
±2V
/±20V/±200V DC header selectable ranges, 3.5 digit.
UM-40-ACA
....AC amps, Scaled or True RMS, (
1 or 5 Amp
internal shunt), 4.0 digit.
UM-40-ACV
....AC volts, Scaled or True RMS.
199.9V AC
/600V AC header selectable, 4.0 digit.
UM-45-DCA
.....DC mV ±20mV/±50mV/
±100mV/
±200mV header selectable ranges, 4.5 digit
UM-45-DCV
....DC Volts
±2V
/±20V/±200V DC Header selectable ranges, 4.5 digit.
UM-35-CL
.......Process 4 to 20mA
(100.0)
, optional 24V DC excitation ,easily user
scalable in engineering units from –1999 to +1999. 3.5 digit
UM-35-HZ
.......
15Hz
to
199.9Hz
or optional 40Hz to 400Hz range up to 500V AC , 3.5 digit.
UM-35-Pressure
..Pressure, strain gage and load cell, 4 and 6 wire, 5V DC excitation,
Header Selectable Sensitivity 2mV/V, 5mV/V, 10mV/V,
20mV/V
, 3.5 digit
UM-35-JF
........
J
thermocouple input, 1° resolution, order
°C
or
°F
, 3.5 digit
UM-35-KF
.......
K
thermocouple input, 1° resolution, order
°C
or
°F
, 3.5 digit
UM-35-RTD/F
..100Ω platinum RTD, 3 or 4 wire, order
°C
or
°F
and
0.1°
or
1°
, 3.5 digit