74
4.8.2 MANU:NAME
Set
Query
Description
Sets or returns the test name for the selected manual test.
The test must be in MANU mode before this command can be
used.
Note only alphanumeric characters (A-Z, a-z, 0-
9) and the “_”
underscore character can be used to set the MANU test
name.
Syntax
Query Syntax
MANU:NAME <string>
MANU:NAME?
Parameter/ Return
parameter
<string> 10 character string. (first character must be a letter)
Example
MANU:NAME test1
Sets the manual test name to “test1”.
4.8.3 MANU:RTIMe
Set
Query
Description
Sets or returns the Ramp Time for the test in seconds.
Note: A “TIME ERR” will result if the Ramp Time + Test Time
is ≥ 240 seconds when the HI SET limit is over 80mA . This
applies to the ACW function only.
Syntax
Query Syntax
MANU:RTIMe <NR2>
MANU:RTIMe?
Parameter/ Return
parameter
<NR2>
0.1~999.9 seconds
Example
MANU:RTIM 0.5
Sets the ramp time to half a second.
4.8.4 MANU:EDIT:MODE
Set
Query
Description
Sets or returns the mode (ACW, DCW, IR) of the selected
manual test.
Syntax
Query Syntax
MANU:EDIT:MODE {ACW|DCW|IR|GB}
MANU:EDIT:MODE?
Parameter/ Return
parameter
ACW
AC Withstand mode
DCW
DC Withstand mode
IR
GB
Insulation Resistance mode
GB mode
Example
MANU:EDIT:MODE ACW
Sets the mode to ACW.