95
I < SET
For GB tests. Current too low. Indicates that the
SOURCE L or SOURCE H test lead is open or poorly
connected. Test the test lead connection with the DUT
to confirm.
I > SET
For GB tests. Current is too high.
R = 0
For GB tests. Resistance = 0. This error indicates that
there is
an error with the measured resistance (0Ω).
Perform the zeroing function again.
6.3 STW-9000 Specifications
The specifications apply when the STW-9000 is powered on for at least 30 minutes
at 15˚C~35˚C.
6.3.1 Specifications
Environment
Range
Temperature
Humidity
Warranty
15˚C ~ 35˚C
≤70% (No condensation)
Operation
0˚C ~ 40˚C
≤70% (No condensation)
Storage
-
10˚C ~ 70˚C
≤85% (No condensation)
Installation Location
Indoors at an amplitude of up to 2000m.
AC Withstanding Voltage
Output Voltage Range
0.100kV~ 5.000kV
Output Voltage Resolution 2V
Output Voltage Accuracy
(1% of s5V) with no load
Maximum Rated Load
(Table1)
STW-9800:
200 VA (5kV/ 40mA)
STW-9900:
500 VA (5kV/100mA)
Maximum Rated Current
STW-9800:
40mA
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 4
0mA(0.5kV<V≤5kV)
STW-9900:
100mA
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 100mA(0.5kV<V≤5kV)
Output Voltage Waveform
Sine wave
Frequency
50 Hz / 60 Hz
Voltage Regulation
1% +5V
[Maximum rated load → no load]
Voltmeter Accuracy
(1% of 5V)
Current Measurement
Range
STW-9800:
0.001mA~40.0mA
STW-9900:
0.001mA~100.0mA
Current Best Resolution
STW-9800:
1uA
0.001mA (0.001mA~0.999mA)
0.01mA
(01.00mA~09.99mA)
0.1mA
(010.0mA~040.0mA)
STW-9900:
1uA
0.001mA (0.001mA
~
1.100mA)
0.01mA
(01.11mA
~
11.00mA)
0.1mA
(011.1mA
~
100.0mA)