6-1. Output HI-R Function
The PS-A power supply unit is equipped with a discharge circuit for removing the electric charge from the output
capacitor in order to maintain an output voltage of 0V while the output is off.
The discharge circuit for the output capacitor does not activate while the output HI-R function of the unit is "ON".
Use the output HI-R function when the unit is employed as a power unit for charging batteries or for electroplating.
6-1-1. Setting the output HI-R function
Operation procedure
1. Press the MENU key while the output is off.
The key is then lit in green.
The ESC/KEY LOCK key and the ENTER/MEM
key then blink green.
The function name is displayed in the voltage
indicator, and the function setting is displayed
(blinking) in the current indicator.
Turn the rotary encoder so that "ON" is displayed
in the current indicator.
The discharge circuit is
activated while the output is
off (HI-R: OFF).
2.
The discharge circuit is NOT
activated while the output is
on (HI-R: ON).
3. Press the green-blinking ENTER/MEM key to
enter the setting.
After the setting is entered, the current indicator
stops blinking, remains lit for a short while, and
then starts blinking again.
4. Press the green-blinking ESC/KEY LOCK key to
exit.
Normal operation of the unit is then available.