11-3-27. Notification setting 2 (AMASK)
This command sets and queries the character string issued for alarm notification 2.
AMASK x1
cation 2.
x1: Uses an 8 digit combination of 0 and 1 to enable/disable
alarm notifi
Setting
AMASK 000000x1 :
AMASK 0000001x :
larm notification.
P occurrence notification.
Enable(1)/disable(0) other a
Enable(1)/disable(0) OC
Application example AMASK 00000001
Sends an alarm notification when OCP occurs.
Query
AMASK?
Queries the alarm notification 2 setting.
Response example
AMASK 00000001
Issues an alarm notification when OCP occurs.
The message issued contains a header to identify the unit on the local bus, followed by the
system address (%1) of the unit on which the alarm occurred, and the alarm type.
The default setting is 00000000.
Remarks
Other alarm
e
ALARM %1,5
OCP occurrenc
ALARM %1,4
11-3-28. Notification setting 3 (SMASK)
This command sets and queries the character strin
s notification.
ble/disable
g issued for sequence statu
SMASK x1
x1: Uses an 8 digit combination of 0 and 1 to ena
sequence status notification.
Setting
SMASK 000000x1 :
SMASK 0000001x :
ble(1)/disable(0) sequence end notification.
Enable(1)/disable(0) sequence pause notification.
Ena
Application example
MASK 00000001
Sends a status notification when a sequence is paused.
S
Query SMASK? Queries
the sequence status notification setting.
Response example
SMASK 00000001
Issues a status notification when a sequence is paused.
The message issued contains a header to identify the unit on the local bus, followed by the
ress (%1) of th
00000
system add
The default setting is 00
e unit on which the alarm occurred, and the alarm type.
0.
Remarks
S
S
equence pause
equence end
PAUSE %1,1
END %1,1
11-3-29. Querying alarm st
This command queries
Query
A
Q
atus (ALM)
alarm status.
LM?
ueries the alarm status.
Response example
A
0
1
LM 1
: Indicates normal operation.
: Indicates an alarm has occurred.
Remarks
11-3
IT
and resets
Setting
-30. Alarm reset (EX _ALM)
This comm
alarms for recoverable conditions.
EXIT_ALM No
parameter
Application example EXIT_ALM
Reset alarm
Remarks
R
re as fo
voltag
e
(115%
output)
F
e powe
off and then on again.
esettable alarms a
llows:
Abnormal sensing
e
Front output terminal ov rcurrent
Abnormal internal heat
Abnormal output
or more of rated
or other alarms, th
r must be tuned
11-3-31. Clearing sequence
LR)
This command clears sequence programs.
Setting SCLR
x1,x2
Clears a sequence program specified within the range of x1, x2.
programs (SC
Application example SCLR 1,10
Clears a sequence program from STEP 1 to STEP 10.
Remarks
x2 must be a number that is equal to or greater than x1.