41
Display
Mode
CC
I Range
H 35A
V Range
H 150V
Function
Dynamic
Configure
0.00
V
0.000
A
0.00
w
Timer1
0.025
SlewRate
2500.00
SlewRate
2500.00
LOAD
RS232
Fine
31/Oct/2013
ms
mA/us
mA/us
Slewrate
settings
Slewrate
settings
2-2-5. CV Mode Response Speed
Description
The response speed setting is the response
speed for the negative feedback control of the
load current when used in CV mode. Response
speed settings are only applicable to CV mode.
A response speed that is too fast could cause
the unit to be unstable.
Reducing the response speed can improve
stability.
Operation
1.
Make sure the load is off.
2.
Press
Main
. Make sure the unit is in CV
mode by using the
Mode[F1]
soft-key.
3.
Select the response speed with the
Response[F4]
soft-key.
Range:
Slow, Fast
Display
Mode
CV
I Range
H 35A
V Range
L 15V
Response
Slow
Configure
0.000
V
0.000
A
0.00
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
31/Oct/2013
V
V
Response
setting