114
4-1-14-2. Current Monitor Output
Description
The voltage output from the IMON OUT terminal
and from the IMON pin on the J1 connector is
used to represent the current input level.
The voltage range used to represent the full
scale current range from the IMON OUT terminal
and from the IMON pin on the J1 connector
depends on the current range settings.
Current
IMON V
Rated
Current
Monitor Output Range
Monitor Connector
Current Range
Monitor
Output
Range
I MON OUT (BNC)
H, L
0 ~ 1V
M
0 ~ 0.1V
I MON (J1)
H, L
0 ~ 10V
M
0 ~ 1V
I MON OUT
BNC
Connector
The IMON OUT BNC connector outputs a
voltage of 0 ~ 1V for the High and Low current
ranges and 0 ~ 0.1V for the Middle current
range. The common potential is connected to
the chassis ground potential.
J1 Connector
The voltage across pins 2 and 3 outputs a
voltage of 0 ~10V for the High and Low current
ranges and 0 ~ 1V for the Middle current range.
The common potential is connected to A COM
(negative load terminal).