148
4-2-2. Connection
Description
The
frame control ports J1 and J2 connectors are used
for control during parallel operation. Up to 5 units can be
used in parallel.
Note
Only the rear panel terminals can be used for parallel
operation, the front panel terminals have a lower current
rating and thus should not be used for parallel operation.
Connection
Cautions
Only the rear terminals can be used for parallel
connections.
Make sure all connections are correct before turning on
the load. Incorrect connections could damage the units.
Only units of the same type and rating can be used in
parallel (except for when the LSG-2100SH booster pack
is used with the LSG-1050H).
Ensure that wiring of sufficient gauge is used when using
parallel connections.
If using
remote
sense, only connect the master to the
voltage sense terminals.
Ferrite
cores
MASTER
J1
J2
+IN
-IN
+S
-S
SLAVE1/
BOOSTER1
J1
J2
+IN
-IN
+S
-S
SLAVE4/
BOOSTER4
J1
J2
+IN
-IN
+S
-S
Load
wiring
Sense
wiring
Frame
link
cable
(GTL-255
)
No ferrite core
to last J1 connecter
+OUT
-OUT
Power
source