35
Table 4-2
Bandwidth
AC Filter
Time
( seconds per reading)
3 Hz – 300 kHz
Slow
7
20 Hz
–
300 kHz
Medium
1
200 Hz
–
300 kHz
Fast
0.1
How to set AC Filter in AC measurements:
User can set the AC Filter either through the front panel operation or through the remote
interface operation.
Front Panel Operation
Press CONFIG + ACV button. Use
◁
and
▷
buttons to locate “BAND WIDTH”
submenu, and then press ENTER to select it. Look for the desired bandwidth using
◁
and
▷
buttons. Then press ENTER on your selection. There are three options: 3Hz,
20Hz and 200Hz. The locations of these buttons are shown with red rectangle frames in
Figure 4-2
.
Figure 4-2
Remote Interface Operation
From your PC terminal, use the following command to specify the filter type:
DETector:BANDwidth {3|20|200|MIN|MAX}
3.1.2.2 Digital Filter
Definition:
DL-2060 uses an averaging digital filter to yield a reading for display from a specified
number of measurement readings in the past. The past measurement readings are
stored in a stack of memory. The number may be in the range of 2 to 100. User may
select one of the two modes of digital filter operations: Moving Average and Repeating
Average.
Summary of Contents for DL-2060
Page 37: ...31 Figure 3 17 1 2 3 5 5 4...
Page 136: ...130...
Page 137: ...131...
Page 138: ...132 D DL 2060 s Dimension...