20
2
Basic Measurement Function
This chapter introduces some basic measurement functions in DL-2060. You will learn
how to use your DL-2060 multimeter to measure voltage, current, resistance, frequency,
period, continuity, diode and temperature in this chapter.
2.1 Voltage Measurements (DC & AC)
The ranges for DC voltage measurements in DL-2060 are 100mV, 1V, 10V, 100V and
1000V. For AC voltage measurements, the ranges are 100mV to 750V AC-Coupled
TRMS
,
or 1000V peak.
Figures 3-1
and
3-2
shows the locations of the buttons needed and
message display for voltage measurement.
Figure 3-3
shows the location of the input
terminals on the rear panel.
Warning: Do not apply more than 1000V (peak) to the multimeter. Applying
excess voltage may damage your meter, or cause possible electric shock or personal
injury.
※
Note: To eliminate the thermal EMFs due to the differences between two metals, use
copper leads to connect your source signal to the meter.
How to measure voltage
①
Selects input signal connections on front or rear panel.
②
Connect the test leads to the terminals as shown in
Figure 3-1
(DC) or
Figure 3-2
(AC).
③
Set RESOLUTION of DCV (Refer to 3.1.3), BAND WIDTH of ACV (Refer to 3.1.2.1) or
skip this step if default is to be used.
④
Press DCV or ACV button for DC or AC voltage measuring.
⑤
Select the auto-range function by pressing AUTO button on the front panel or use
△
▽
buttons to select desired range.
⑥
Connect test leads to your source signal and observe the reading shown on the
display. If the input signal is beyond the allowed range, an overflow message “OVLD”
will be displayed.
Summary of Contents for DL-2060