126
*SRE?
Query the Status Byte enable register. The multimeter returns a decimal value which corresponds
to the binary-weighted sum of all bits set in the register.
*STB?
Query the Status Byte summary register. The *STB? command is similar to a serial poll but it is
processed like any other instrument command. The *STB? command returns the same result as
a serial poll but the “request service” bit (bit 6) is not cleared if a serial poll has occurred.
B.10 SCPI Compliance Information
This section encloses a list of commands that are device-specific to the DL-2060.
Although not included in the 1999.0 version of the SCPI standard, these commands are
designed with the SCPI format in mind and they follow all of the syntax rules of the
standard.
Many of the required SCPI commands are accepted by the multimeter but are not
described in this manual for simplicity or clarity. Most of these non-documented
commands duplicate the functionality of a command already described in this chapter.
MEASure:
CONTinuity?
DIODe?
SAMPle:
COUNt {<
value
>|MINimum|MAXimum}
COUNt? [MINimum|MAXimum]
[SENSe:]
FUNCtion "CONTinuity"
FUNCtion "DIODe"
FREQuency:VOLTage:RANGe {<
range
>|MINimum|MAXimum}
FREQuency:VOLTage:RANGe? [MINimum|MAXimum]
FREQuency:VOLTage:RANGe:AUTO {OFF|ON}
FREQuency:VOLTage:RANGe:AUTO?
PERiod:VOLTage:RANGe {<
range
>|MINimum|MAXimum}
PERiod:VOLTage:RANGe? [MINimum|MAXimum]
PERiod:VOLTage:RANGe:AUTO {OFF|ON}
PERiod:VOLTage:RANGe:AUTO?
ZERO:AUTO?
Summary of Contents for DL-2060
Page 37: ...31 Figure 3 17 1 2 3 5 5 4...
Page 136: ...130...
Page 137: ...131...
Page 138: ...132 D DL 2060 s Dimension...