B.4 The CONFigure Command
CONFigure:VOLTage:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC voltage measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:VOLTage:DC:RATio {<range>|MIN|MAX|DEF },{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC:DC ratio measurements with the specified range and
resolution. This command does not initiate the measurement. The specified range applies to the
source signal and autorange is selected for the reference signal.
CONFigure:VOLTage:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for AC voltage measurements with the specified range and
resolution. This command does not initiate the measurement. For AC measurement, resolution is
fixed at 6½ digits. Therefore the resolution parameter only affects the front panel display.
CONFigure:CURRent:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC current measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:CURRent:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for AC current measurements with the specified range and
resolution. This command does not initiate the measurement. For AC measurement, resolution is
fixed at 6½ digits. Therefore the resolution parameter only affects the front panel display.
CONFigure:RESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 2-wire Ω measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:FRESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 4-wire Ω measurements with the specified range and
resolution. This command does not initiate the measurement.
CONFigure:FREQuency {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for frequency measurements with the specified range and
resolution. This command does not initiate the measurement. For frequency measurements, the
meter uses only one “range” for all inputs between 3Hz and 300kHz. With no input signal applied,
frequency measurements return “0”.