14
1. Activate continuity test Press the
key twice.
2. Continuity mode
display appears
+ Ω
Indicates Continuity test
CONT
2nd display shows the title
3. Connect the test lead
and measure
Connect the test lead between the Ω and the COM
port. The display updates the reading.
COM
2-15. Set continuity threshold
Background
Continuity threshold defines the maximum resistance allowed in the DUT when
testing the continuity.
Threshold Range
0
Ω~ 1000Ω, 1Ω resolution, 10Ω default
1. Activate threshold
setting
1. Press the Shift key, the 2ND
key, the Right key. The
measurement menu appears.
2.
Press the Down key, the Right
key, the Enter key. The continuity
threshold setting appears.
2. Edit threshold
3.
Move the cursor (the flashing digit) using the
Left/Right key.
4.
Change the value using the Up/Down key.
Range:
0
Ω~ 1000Ω, 1Ω resolution, default 10Ω
3. Go back to the
default display
Press the Enter key to confirm the edited
threshold. Press the Exit key to go back to
the default display.