13
2W connection
COM
4W connection
COM
4W
SENSE
LO
2-12. Select Resistance range
Auto range
To turn the automatic range selection On/Off, press
the AUTO key.
Manual range
Press the Up or the Down key to select the range.
AUTO indicator turns Off automatically. If the range is
unknown, select the highest range.
Selection list
Range
Full scale @ slow rate
100
Ω
120.000Ω
1k
Ω
1.20000kΩ
10k
Ω
12.0000kΩ
100k
Ω
120.000kΩ
1M
Ω
1.20000MΩ
10M
Ω
12.0000MΩ
100M
Ω
120.000MΩ
Note
For more detailed range, see the specifications at page51.
2-13. Diode Test
Background
Diode test checks the forward bias characteristics of a diode by running a
constant forward bias current, approx. 0.5mA, through the DUT.
1. Activate diode test
Press the
key once.
2. Diode mode display
appears
+ V
Indicates Diode test
DIODE
2nd display shows the title
3. Connect the test lead
and measure
Connect the test lead between the
and
COM port; Anode-V, Cathode-COM. The display
updates the reading.
COM
2-14. Continuity Test
Background
Continuity test checks that the resistance in the DUT is low enough to be
considered continuous (of conductive nature).